IsoCube

IsoCube Isokinetic Wireless Module

Description

The IsoCube module from AMS Analitica represents an important innovation for flow measures and for stack sampling, effectively simplifying procedures thanks to the innovative wireless connection with the AirCube Touch series samplers.

The system consists of an measuring module attached directly to the probe that allows the elimination of long cables and connecting pipes, limiting the connection between the probe and the sampler to the air intake pipe only. Communication with the sampler takes place thanks to a Bluetooth connection, which once stable will activate a software dedicated to performing the sampling of pollutants in emission compliant with US EPA and CEN requirements in automatic isokinetic mode.

The IsoCube is able to detect differential pressure, absolute pressure and static pressure, it is able to measure the Swirl angle thanks to the triaxial transducer inserted inside the module and has four inputs for stack temperature detection or for other use.

Main features

  • Can be connected directly to the isokinetic probe allowing the elimination of long cables and connecting tubes
  • Elimination of interference in temperature and differential pressure sensors due to cable length
  • Automatic chimney calculation of the turbulence correction factor (Swirl angle)
  • Three pressure sensors: differential, static and barometric for the calculation of absolute pressure in accordance with ISO 16911:2013 requirements
  • Four entrances for temperature detection
  • No serial cable connection between IsoCube and sampler
  • Parameters controlled directly by the sampling unit, effectively transforming it into an automatic isokinetic model sampler

Technical features of the IsoCube Meter Module

  • 4 inputs for temperatures (in-stack, impinger, condensate and filter holder) up to 1100°C – Ris. 0.01°C set by choice between thermocouple of type K, J, S
  • Atmospheric barometric pressure sensor range 500-1100 mbar (RES. 0.01 mbar)
  • Triaxial sensor for digital SWIRL angle measurement (x, y, z)
  • Static pressure sensor

    • Measurement range (FS): +/- 1000 mm/H2O
    • Resolution: 0.1 mm/H2O

    Accuracy:

    • 0.9% on scale (FS) 0..100 mmH2O with readings less than 010 mmH2O
    • 0.8% on scale (FS) 0.300 mmH2O with readings lower than 030 mmH2O
    • 0.8% on scale (FS) 0.700 mmH2O with readings lower than 070 mmH2O

  • Differential pressure sensor:

    • Measurement range (FS): 0-100 mm/H2O
    • Resolution: 0.01 mm/H2O

    Accuracy:

    • 1% on scale (FS) 0..20 mmH2O with readings less than 010 mmH2O
    • 0.5% on scale (FS) 0.50 mmH2O with readings lower than 030 mmH2O
    • 0.3% on scale (FS) 0..100 mmH2O with readings less than 070 mmH2O

  • Environmental conditions of use

    • Temperature -10°C % +50°C
    • Relative humidity 95% non-condensing

Aircube com2-th touch sampler technical features in combination with the isocube isokinetic module

Calculation of isokinetic parameters in accordance with ISO-CEN 16911, US EPA M1, M2and M3

Calculation of the flow rate for sampling PM10 and PM2.5 according to US-EPA 201A and EN-23210

Automatic calculation of traverse points according to EN-13284:2017 ISO-CEN 16911, EN 15259, US EPA M1 and M2

Selection of sampling nozzles in automatic and manual mode

Isokinetic protocol in accordance with ISO-9096, EN-13284 and US EPA

Automatic angle calculation of SWIRL with three-axis transducer, automatically calculated flow and constant without adding data in manual mode

Automatic calculation of turbulence (WAF) resulting from the chimney wall effect (EPA 2H)

Step-by-step settings of sampling parameters in accordance with European and US EPA regulations

Graphic visualization of stack diagram and real-time sampling points

Single Diaphragm suction pump

Suction flow rate between 0.4 and 30 liters per minute with automatic back pressure compensation

4 inputs for the detection of temperatures (stack, impinger, condensate and filter holder) up to 1100°C – Ris. 0.01°C set by choice between thermocouple of type K, J, S

Flow control with volumetric meter and digital encoding system

Color graphic display Touch Screen 4.3”

Atmospheric barometric pressure sensor range 500-1100 mbar (RES. 0.01 mbar)

Static pressure sensor

– Measurement range (FS): +/- 1000 mm/H2O

– Resolution: 0.1 mm/H2O

– Accuracy:

– 0.9% on scale (FS) 0..100 mmH2O with readings less than 010 mmH2O

– 0.8% on scale (FS) 0.300 mmH2O with readings lower than 030 mmH2O

– 0.8% on scale (FS) 0.700 mmH2O with readings below 070 mmH2O

Sensor for differential pressure:

– Measurement range (FS): 0-100 mm/H2O

– Resolution: 0.01 mm/H2O

– Accuracy:

– 1% on scale (FS) 0..20 mmH2O with readings less than 010 mmH2O

– 0.5% on scale (FS) 0.50 mmH2O with readings lower than 030 mmH2O

– 0.3% on scale (FS) 0..100 mmH2O with readings less than 070 mmH2O

Environmental conditions of use

Temperature -10°C % +50°C

Relative humidity 95% non-condensing

Communication interfaces Bluetooth, USB

Internal memory with SD Card: 32 Gb standard option

Dimensions 28x28x30 cm (WxDxH) – Weight 8 Kg

220 V-AC power supply and integrable external battery pack (Optional)

Request information